Introduction

The medical practice of measuring blood pressure by constricting the upper arm and listening with a stethoscope to assess when the first sound and final sounds are audible dates back over 100 years ago. This method is still considered the gold standard for noninvasive blood pressure measurement. While there have been advancements in the technique for upper arm constriction, the quality of stethoscopes, and most recently fully automatic blood pressure devices – the overall process has not changed significantly. The discovery and utilization of Korotkoff sounds (KorS) are what allows physicians to check patient blood pressures and provide appropriate medical treatments. Understanding the underlying physiology and proper measurement techniques are important for quality patient care and appropriate medical therapy.

Issues of Concern

Korotkoff sounds (KorS) are the audible noises used to measure blood pressure. Throughout the short process of taking a blood pressure measurement, the sounds change a phenomenon which has undergone extensive study. Korotkoff sounds are broken down into five distinct phases, each with a distinct sound and waveform associated with the phase. Phase I has a clear tapping tone. Phase II is associated with a softening of the tapping and a swishing element. Phase III sounds like phase I but with distinct sharpening. Phase IV is noted to have abrupt muffling of the sounds, followed by Phase V which is the cessation of all sounds. When measuring blood pressure, a cuff is applied to the upper arm and inflated to the point that blood is no longer able to flow through the brachial artery. When pressure is released to point blood is again able to flow, phase I begins. As pressure continues to decrease the examiner can hear the different phases, ultimately concluding with phase V when no sound is audible. The first sound appears as the pressure approaches systolic blood pressure (SBP) and the diastolic blood pressure (DBP) measurement is by the disappearance of sound.[1] Despite 100 years of debate and research, the underlying physiology of Korotkoff sounds is not fully understood. There are several theories that are prominent in the literature trying to explain the underlying cause of Korotkoff sounds. There are two main schools of thought guiding the research of Korotkoff sounds; the sounds are a result of fluid turbulence or the result of arterial wall oscillations. Adding a layer of complexity, those who test the arterial walls are uncertain what the best method of experimentation is. For example, Dr. Charles Babbs used a primarily mathematical approach to analyze wall motion mechanics, while Drzewiecki et al. used an electrical analogy in attempts to postulate the equations governing the fluid dynamics and nonlinear compliance of arterial segments. While there are many ideas, two commonly speculated theories for the causation of the sounds heard in stethoscopes will be briefly discussed here.[2]

Dr. Nikolai Korotkov is the physician credited with discovering the sounds that carry his name. He speculated the sounds resulted from the closing and opening of the vessel. He suspected the net transmural pressure oscillating from positive to negative was responsible for the different phases of sound. Dr. Charles Babbs, conducted an experiment measuring blood pressure and the associated Korotkoff sounds, concluding the sounds were from the transition of the vessel from a buckled to fully expanded the state. The transient vibrations of the artery being the mechanism for actual sound production. The study, while not universally accepted as proof for the physiologic source of Korotkoff sounds. It was able to explain how venous engorgement downstream to the collapsed artery adds to the increased resistance, preventing artery collapse, as the cause of the muffling heard in phase IV.[2]

A study performed at Chengdu University in China demonstrated the importance of proper technique when measuring blood pressure (BP). The study concluded that different stethoscope positioning results in sounds being present sooner/later, resulting in higher/lower BP readings than the actual blood pressure. Over the last 100 years, it has been widely accepted that hearing the first sound is in direct correlation with SBP. However, there is still discourse regarding the measurement of the DBP. It has actually become known as the “diastolic dilemma” since the argument is whether the DBP should be measured during phase IV when the sound muffles or in phase V when the sound disappears completely. The position of the stethoscope also affects auscultation and influences the DBP reading by up to 10 mm Hg. The study was able to recommend stethoscope positioning which most accurately represented the DBP, via comparison to an invasive blood pressure measurement simultaneously. The study concluded the proper placement of the cuff, just above the elbow, with the stethoscope under the cuff provided the most accurate blood pressure readings. This study along with many others have demonstrated how a seemingly simple procedure can have extensive variability, dependent on multiple factors (cuff pressure, circumferential and longitudinal stethoscope positioning, and sound cutoff). Ongoing research and continued technological advancements will hopefully definitely pinpoint the true physiology responsible for Korotkoff sounds, enabling even more precise blood pressure readings and patient care.[2][1][3][4][5]

Clinical Significance

As medical technology continues to advance, machines are assuming on a more significant role in everyday patient care. One of the most prominent automated medical tests is blood pressure readings, whether in the clinic, the ICU, or labor deck. Patients are having their blood pressure (BP) measured using devices which rely on oscillometric methods to give medical providers information to guide treatment. Extensive data analyses of these machines have demonstrated an overestimation of hypotensive BPs and an underestimation of hypertensive BPs. The reading of BP within normal ranges have proven to be accurate and reliable.[6] However, in most scenarios, it is the abnormal BPs that are most important to medical providers. It is quintessential for accurate blood pressure readings in stroke, acute kidney injury, septic, etc., patients. Blood pressure is easy to assess and easy to correct to prevent further deterioration. A wise clinician will often advise staff to manually check the BP if the machine reading is suspicious compared to patient status.[7] Despite the advancement of technology, we still rely on human measurement for a second and more trusted blood pressure reading. It is for this reason, understanding KorS and how to properly obtain BP readings is crucial. Accurate blood pressure readings obtained manually via cuff and stethoscope are still a vital aspect of modern medicine, and with a greater understanding of Korotkoff sounds, we will more precisely measure BP and allow better clinical decision making.


References

[1] Pan F,Chen F,Liu C,Yang Z,Liu Z,Zheng D, Quantitative Comparison of Korotkoff Sound Waveform Characteristics: Effects of Static Cuff Pressures and Stethoscope Positions. Annals of biomedical engineering. 2018 Nov;     [PubMed PMID: 29959551]
[2] Babbs CF, The origin of Korotkoff sounds and the accuracy of auscultatory blood pressure measurements. Journal of the American Society of Hypertension : JASH. 2015 Dec;     [PubMed PMID: 26553392]
[3] Liu C,Griffiths C,Murray A,Zheng D, Comparison of stethoscope bell and diaphragm, and of stethoscope tube length, for clinical blood pressure measurement. Blood pressure monitoring. 2016 Jun     [PubMed PMID: 26741415]
[4] Xiang H,Liu Y,Qin Y,Cao Z,Guo T,Yu M, A pilot application of Korotkoff sound delay time in evaluating cardiovascular status. Technology and health care : official journal of the European Society for Engineering and Medicine. 2015     [PubMed PMID: 26410508]
[5] Trigg SA,Abreu D,Bitton-Foronda B,Foley FC,Gibson AL, Comparison of Systolic Blood Pressure Measurements by Auscultation and Visual Manometer Needle Jump. International journal of exercise science. 2019     [PubMed PMID: 30761196]
[6] Meidert AS,Saugel B, Techniques for Non-Invasive Monitoring of Arterial Blood Pressure. Frontiers in medicine. 2017;     [PubMed PMID: 29359130]
[7] Xiang H,Liu Y,Li Y,Qin Y,Yu M, Noninvasive measurement of beat-to-beat arterial blood pressure by the Korotkoff sound delay time. Blood pressure monitoring. 2012 Feb     [PubMed PMID: 22190147]